IK 5000 QUADRA-CHEK 45
1.1 Operation
Enable or prohibit data shifts for Profile3D fit
The Profile3D fit algorithm shifts the probed data points in the
cartesian or polar coordinate system to achieve the best fit between
the probed points and the nominal part profile.
Maximum degrees of freedom are given to the fit algorithm by
checking the Allow X, Allow Y, Allow Z, Allow AX, Allow AY and
Allow AZ check boxes across the bottom of the window. Clearing a
check box prohibits data movement in the indicated orientation.
To give maximum degrees of freedom to the Profile3D fit algorithm:
Check the Allow X, Allow Y, Allow Z, Allow AX, Allow AY and
Allow AZ check boxes across the bottom of the Measure Profile3D
window
Clear a check box to prohibit movement of data in the indicated
orientation
Check all check boxes for maximum degrees of freedom
Allowing the maximum degrees of freedom for the
Profile3D fit algorithm is recommended for most
applications.